EasyManua.ls Logo

GRAPHTEC GL450 - Logic;Alarm Functions

GRAPHTEC GL450
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Checks and Preparation
2-17
2.14 Logic/Alarm Functions
Connect the round connector of the logic alarm cable (B-513, option) to the logic Input/alarm output terminal
on the GL450.
Logic alarm cable
(B-513: 2m)
Logic Functions
Alarm Functions
Alarm Output Circuit
Note: Be sure to not to exceed the maximum ratings
Maximum rating
VCEO (voltage between connector and emitter): 30V
IC (connector current): 0.5A
+5V
100K
GL450
Wiring
The cables have bare tips. Please perform wiring as required.
Item Description
Number of channels 4
Input voltage range 0 to +24V max. (single-ended ground input)
Threshold level +2.5V
Hysteresis Approx. 1 V (+2 to +3 V)
Item Description
Number of channels 4
Maximum rating VCEO (voltage between connector and emitter): 30V
IC (connector current): 0.5A
Item Number Lead Color
Logic signal 1 Orange/red
2 Orange/black
3 Gray/red
4 Gray/black
Alarm output 1 White/red
2 White/black
3Yellow/red
4Yellow/black
Common ground GND Pink
GND Pink
GND Shielded lead
Orange/red
Orange/black
Logic signals
Gray/red
Gray/black
White/red
White/black
Alarm output
Yellow/red
Yellow/black
Pink
Pink
GND
Shielded lead
Logic alarm cable (B-513)

Table of Contents

Other manuals for GRAPHTEC GL450

Related product manuals