3-12
Settings and Measurement
Pulse................................... The signals that can be input to the pulse input terminal for data
processing are Revol., Counts, and Inst.
Modes: Select from OFF, Revol., Counts, and Inst.
• Off: Input is disabled.
• Revol.: Counts the number of pulses per second, and displays the
values multiplied by 60 as rpm values.
• Counts: Displays the cumulative number of pulses for each
sampling interval from the start of measurement.
• Inst.: Displays the number of pulses for each sampling interval.
Range: Can be set for each of the modes.
EU: Scales and converts the measured value.
• Meas. Value:Specifies the numeric value to be converted.
• EU Value: Specifies output after conversion.
• Unit: Selects the converted unit, which can be specified as
a user-defined character string consisting of
alphanumerics. The Unit parameter can also be
specified by selecting the Select Unit setting.
• Select: Selects the type of engineering unit.
• Choose: Selects the converted unit. The Unit displayed here is
the type of unit selected by the Select Unit setting.
Slope: Sets the condition for the input signal operation.
• H: Operates when the signal is a rising signal
• L: Operates when the signal is a falling signal.
Logic ................................... Enables logic amps to be used: Off (disabled), On (enabled).
Color specification for each channel:
This parameter enables the automatically set value for each channel
to be changed manually.
Mode Range Maximum number
of pulse inputs
Revol. 500, 5k, 50k, 500k Revol./F.S. 50k/s
Counts 50k, 500k, 5M, 50M, 500M C/F.S. 50k/sampling interval
Inst. 50k, 500k, 5M, 50M C/F.S. 50k/sampling interval