EasyManua.ls Logo

GRAPHTEC GL450 - Page 49

GRAPHTEC GL450
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-14
Settings and Measurement
ee
ee
e DATA Settings Window
Data capture, Data replay, and Statistical Calculation settings are made here.
DATA Menu Structure
Record Settings
Detailed settings for data capture are performed in this section.
Sampling Interval ................ Specifies the sampling interval for data capture. 16 intervals are
provided.
CHECKPOINT
The maximum sampling interval that can be selected depends on the
number of channels.
• 10ch: 100 ms • 30/40/50 ch:500ms
• 20ch: 200 ms • 60ch or more: 1 sec
Capture Destination ............ Selects the destination for saving measured data.
Memory: 4 Mbytes (2 Mwords)
This is the total amount of memory. If the number of
channels used is increased, the amount of data that
can be captured is reduced.
Setting Selections available Setting method
Record Settings
100, 200, 500ms 1, 2, 5, 10, 20, 30s ENTERSelectENTER
• Sampling Interval 1, 2, 5, 10, 20, 30min, 1h
Capture Destination Memory, PC Card ENTERSelectENTER
File Name Specify file (only when PC card has been selected)
• File Type: GBD, CSV ENTERSelectENTER
• Name Type: Auto, User ENTERSelectENTER
• File: Folder name, file name ENTERSpecify fileOK
Auto Save Off, I hour, 3 hours, 6 hours, 12 hours, 24 hours ENTERSelectENTER
(only when internal memory has been selected)
Statistical Calculation
Off, Average, Max, Min, Peak, RMS ENTERSelectENTER
• Function
Calculation channel Select the number of channels from the input terminal unit: ENTERSelectENTER
specification
2 ch

Table of Contents