3-94
CHAPTER 3 Settings and Measurement
(9)-5 Cursor Sync
Sets up the function that moves two cursors in synchronization.
Selection item Description
Off Cursors are not synchronized. Only the specied one cursor moves.
On Two cursors move in synchronization. Cursor A is always the fulcrum.
* Cursor Synch is turned Off when you move a cursor using Move to Selected Position or perform Data Search.
(9)-6 Date Search
Sets the search conditions to be used in the next sections (9)-7 "Find Next" and (9)-8 "Find Previous").
The operation is Edge operation.
Selection item Description
CH Sets the channel to be used for search.
CH1-200, GS1-8, WL1-1 to WL5-8 : The specied analog channel is used for search.
Logic1-28 : The specied logic channel is used for search.
Pulse1-28 : The specied pulse channel is used for search.
Alarm1-10, WL1-5 : The specied alarm output is used for search.
Mode Sets the search mode.
Both : Detects an edge at which alarm output changes from generation to cancellation
or vice versa when Alarm is selected.
↑
H : Detects a rising edge of an analog signal or an edge at which alarm output
changes from cancellation to generation
↓
L : Detects a falling edge of an analog signal or an edge at which alarm output
changes from generation to cancellation.
Level Sets a voltage level to be searched for when the search channel is an analog or pulse
channel.
(9)-7 Find Next
Executing this option moves the cursor to a next position where the search conditions are met, down from
the current cursor position. (Specify the search conditions as described in (9)-6 "Data Search.")
(9)-8 Find Previous
Executing this option moves the cursor to a previous position where the search conditions are met, up from
the current cursor position. (Specify the search conditions as described in (9)-6 "Data Search.")