CHAPTER 3 Settings and Measurement
(1)-5 EU (Scaling settings)
This is used to convert the measured signals to other units.
(1)
(2) (3) (4)
(5) (6)
(7)
<For voltage input>
(1)
(8)
<For temperature input>
(2) (3)
Setting Description
(1) EU Function Sets the scaling function to ON or OFF.
(2) Meas. Value (Upper/Lower) Sets the upper and lower limits of values to be converted.
* For temperature input, there is no distinction between upper and lower limits. See the
setting examples shown below for details.
(3) EU Output Value (Upper/
Lower)
Sets the upper/lower limit output values after conversion.
* For temperature input, there is no distinction between upper and lower limits. See the
setting examples shown below for details.
(4) Dec pt Sets the decimal point position for an EU output value.
(5) Select Selects a specic engineering unit classication. (The following are available.)
Cur, Length, Area, Volume, Velocity, Accel, Freq, Mass, Energy, Pressure, Flow , Temp,
Strain, Brightness,Concentration
(6) Unit Selects a unit to be used after conversion. A unit displayed in this eld belongs to the
classication selected in "Select." To set a unit not displayed in this eld, set arbitrary text in
"Arbitrary unit." The setting selected in this eld is displayed in "Arbitrary unit."
(7) Arbitrary Unit Sets a unit to be used after conversion.
Arbitrary text consisting of alphabetical characters and numerical values can be set as a unit.
(For the character input, refer to “(9) Character string input” in “3.4 Setting Menus”.)
When “Select Unit” or “Unit” is specied, it is reected in here.
(8) Reads the current temperature
measurement value
Substitutes the current measurement value into (2) Measurement value and (3) EU output
value.
* The value is not substituted when burnout occurs or the scale is exceeded.
•
If a message window opens, follow the instruction in the message to change the setting value.
•
The Scaling function performs calculation using a ratio of the Meas. Value and EU Output Value settings.
•
The digital display shows “++++/----” when the converted value cannot be processed by the GL240.
•
The span may be changed depending on the Scaling settings.
•
For temperature input, the offset setting for an input value is used.
Meas. Value
+5.000Upper limit
-5.000Lower limit
EU Output Value
+20.00
-20.00
Dec pt
+ xx.xx
Choose
rpm
+5 V
CH.1 10V
-5 V
+ 20.00 rpm
CH.1 Scaling 1
- 20.00 rpm
Setting example: For voltage input
Meas. Value
3°C is always added to the measurement value
22.0℃
EU Output Value
25.0℃
Setting example: For temperature input