3-26
CHAPTER 3 Settings and Measurement
(1)-6 Misc.
(1)
(2)
(6)
(7)
(3)
(4)
(8)
(9)
(5)
The information varies depending on GL240 amplier and GS sensor or terminal/adapter (settable when it is
connected through the wireless sensor).
* Please check the setting target and the information of the item number in the table below.
Setting object Setting Description
Voltage,
humidity,
acceleration
settings
(1) Annotation Settings Set the annotation (comment) displayed in the CH.
(2) Inter-CH Op Settings Set what to do in calculation between channels.
Four arithmetic operations (+, -, x, ÷) can be set as calculation between
channels.
(3) Span Settings Set the upper and lower limits of values of a span in which a waveform
should be displayed.
(4) Waveform Color Setting 0 to 31 for each of red, green, blue (RGB)
(5) Line Thickness Setting Setting 1 to 8 dots
(6) Trace This is used to set the waveform display.
(7) Perform Auto Zero ADJ.
(Auto balance in the case of
acceleration)
The current input voltage (acceleration value) is calculated as a zero
point voltage value (acceleration value).
Automatic adjustable voltage range is within ±10% of set range.
(8) Reset Auto Zero ADJ. Reset the zero point voltage value (acceleration value).
* When the temperature is set, this function is not available.
(9) Set Zero Point as: The zero point voltage value (acceleration value) is displayed.
* When the temperature is set, this function is not available.
When GS sensor and terminal/addapter are used
*
For the number in the setting items in the following table, refer to the voltage, humidity and acceleration
settings described above.
Setting object Setting Description
Temperature
setting
(1), (2), (3), (4), (5), (6) Same content as the voltage.
Accumulated reference
temperature
Set the reference temperature accumulated within the range of -20 to
+85°C for the temperature and humidity sensor.
Accumulated polarity Set High/Low accumulated polarity.
Logic setting (5), (6) Same content as the voltage.
Pulse setting (4), (5), (6) Same content as the voltage.
Number of pulses per revolution Set the number of pulses per revolution.
CO2 setting (1), (2), (3), (4), (5), (6) Same content as the voltage.
CO2 calibration Set the CO2 calibration.
AC clamp setting (1), (2), (3), (4), (5), (6) Same content as the voltage.
Voltage (Power only) Set the measurement voltage value to calculate the power within the
range of 90 to 264V.
Power factor (Power only) Set the power factor to calculate the power within the range of 0.30 to 1.00.
Auto zero adjustment Display the input voltage after releasing the zero point voltage value.
Zero adjustment reset Display the zero point voltage value. (Display only)