3-37
CHAPTER 3 Settings and Measurement
Illuminance / Ultraviolet Sensor of the wireless sensor setting: optional
(GL100-WL+GS-LXUV)
Perform the settings for the illumination / ultraviolet sensor input of wireless sensors.
When the wireless sensor and illumination / ultraviolet sensor are connected, the setting screen is displayed.
(1)-34 Input setting
This is used to select the input conditions.
Setting Description
Off The input signal is not measured. The waveform and digital display also are not performed.
1ch ILLUM. Used to measure the illuminance
2ch UV Used to measure the ultraviolet.
3ch ACC. ILLUM. Used to measure the accumulated illuminance.
4ch ACC. UV. Used to measure the accumulated ultraviolet.
5ch to 8ch No AMP When connecting the two GS series and terminal / addapter using a branch adapter for GS, the
following combination is displayed.
Combination CH1 to 4 CH5 to 8
TH+LXUV TH LXUV
TH+CO2 TH CO2
LXUV+CO2 LXUV CO2
Symbols in the table are as follows:
TH: Temperature and humidity sensor
LXUV: Illumination / ultraviolet sensor
CO2: CO2 sensor
(1)-35 Range setting
This is used to select the illuminance range to be measured. The ultraviolet and accumulated ultraviolet ranges
are xed.
Setting Description
ILLUM. 2000, 20000lx, 200klx (operates in conjunction with 3ch)
UV 30mW/cm
2
xed range
ACC. ILLUM. 2000, 20000lx, 200klx (operates in conjunction with 1ch)
ACC. UV. 30mW/cm
2
xed range
Do not change the range through the GL100-WL after setting the range of the GL240.
The displayed values of the GL240 may not match the measured values.