EasyManua.ls Logo

GREEN INSTRUMENTS G36p - Figure 6-1: Calibration Menu Structure and Quick Guide

GREEN INSTRUMENTS G36p
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
G36p
Oxygen Analyzer
Version 2.11 Revision March 2022 35
+
/
--
+
/
--
Figure 6-1: Calibration menu structure and quick guide
For the span calibration gas, clean dry instrument air with oxygen content of 20.9 % is
recommended.
Before starting to calibrate the analyzer, check the settings of zero and span gas used in
the Calibration Menu.
If the value of zero gas used is not correct, select Zero gas used and enter the value of
the zero gas used.
If the value of span gas used is not correct, select Span gas used, and enter the value
of the span gas used.
Press to go back to the Main Menu
To calibrate the oxygen signal, do the following steps:
Introduce zero calibration gas to the sensor and wait for the zero-value reading to sta-
bilize, e.g. 1.8 %. Then enter Calibration Menu and select Zero calibration. The zero
calibration is completed when the analyzer returns to the Main Menu. The new cali-
bration value will be displayed after 15 seconds (as per default setting of freeze out-
put).
Introduce the span calibration gas to the sensor, and wait for the span value reading to
stabilize, e.g. 21.1 %. Enter Calibration Menu and select Span calibration. The span

Table of Contents

Related product manuals