Routine Maintenance
38
7.4 Gas Connection
The calibration and sample gas connections to the sensor should always be checked and
ensured no leaking. A leaking connection might result in inaccurate measuring results.
To have accurate measuring results, the gasses supplied to the sensor must have a con-
stant flow and pressure. The pressure and flow rate of the sample gas should be the same
as the pressure and flow rate of the calibration gasses.