and the input range of the remote system
→Check and correct output signal settings, see section 5.4
No trend curves
Graph Menu may be
frozen
Y-limits is set too low → reset Y-limits, see section 5.1
X-timing is set too high → restart the analyzer, reset X-timing,
see section 5.1
→Calibrate the touch panel, see section 4.5
No data log
No historical trend
curve
→Check if the SD card is in position
→If yes, check the SD card’s lock mode. The lock button should
always be in the unlock position.
→Replace the SD card if it is defect
Unable to reload de-
fault settings
→ If the screen shows Error!! No SD card, please check the SD
card inside the analyzer. Replace the SD card if it is defect. Con-
tact Green Instruments for a new SD card with the default setting
file
→ If the screen shows Error!! No parm file, please Contact Green
Instruments to resend the default setting file. Copy the
g1parm.bin to the SD card:\bin, then try to reload default settings
one more time.