910FS Splicer
Greenlee / A Textron Company 4455 Boeing Dr. • Rockford, IL 61109-2988 USA • 815-397-7070
27
Maintenance
The 910FS can perform routine maintenance. In “Maintenance”, the operator can
calibrate arc, set motor drive, calibrate motors, adjust screen, replace electrodes, and
stabilize electrodes.
Arc Calibration
Used to calibrate arc function to obtain the best splicing results when a splicing session
is rst started or if the splicing environment has changed.
1. Place splicer in Splice Mode #1 (AUTO).
2. Select “Arc Calibration.”
3. Prepare and load bers.
4. Press
8
.
5. If too low or high, load new bers.
6. Press
8
.
7. Repeat steps 1–4 until “arc adjust OK” displays on splicer screen.