SENSORMOUNTINGLOCATION
-TankLevel/InventoryApplications
SENSORMOUNTING
Eachsensorisequippedwitha¾inch"isolationcoupling"whichMUSTbeusedinyourinstallation.A
threadednippleorlengthofconduitmaybeusedtopositionthesensoratthedesiredheight.
Thesensorshouldbehand-tightened(likealightbulb)byturningthesensorstemonly.DONOTusea
wrenchanddonotovertighten.DONOTclampthesensorbelowtheisolationcoupling.
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 19
SENSORFACE MUST
BE HORIZONTAL ±1°
1FT (30cm) HORIZONTAL
FROM SIDEWALL FOR
EVERY 10' (3m) VERTICAL
NOTE: 2FT (60cm) FOR
EVERY 10' (3m) DEPTHON
APPLICATIONS WITHROUGH
SIDEWALLS, LADDERS,
REINFORCING RINGS ETC.
MAX. LIQUIDLEVEL
PREFERREDSENSOR
LOCATION
6" SCHEDULE 40
STANDPIPE IF
EXTENSION
REQUIRED