RUN
Ascrollingdisplayshowstheunitsselectedfromthe
UNITS/MODE column,themodeofoperation(Range, Level
or Flow),thefullscalevalueforthelargenumericdisplay,
Totalizervalue(in FLOW mode),andairtemperatureatthesensor
location.
LEVELREADING
FromRUNuse È toviewthelevelmeasuredbytheinstrument.
ECHOCONFIDENCE
EchoStrengthisautomaticallyadjustedbytheSLT32inresponseto
operatingconditionsintheapplications.
FromRUNuse È togettotheEchoConfidencedisplay, EC.
TOTALIZER
TheTotalizerdisplaywillonlybeenabledwhentheSLT32iscalibratedin FLOW mode.From RUN use
È or Ç keystodisplay Tot: value.TheTotalizervalueisupdatedevery2secondswithflowvolume
1litre(0.264USG).Thedisplaywillshowupto10digitsandthenrestartat0automatically.
Press È or Ç keystoreturnto RUN.
Thetotalizercanberesettozerobygoing Æ to SPECIALFUNCTION and È to ResetTot?.
RELAYSTATUSDISPLAY
Press È from Tot: to RELAYS1,2,3.
Energizedrelayswillbedisplayedinreversetype RELAYS:
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 7
Current
Echo Strength
Arrow
Minimum
Marker
EC 9
T: 9999999999
T: 0283168470
TOT: 0
TOT: 0
Units: ft³, USG, IG, L
Units = m³
PASSWORD: 00
RUN
Level: 2.000 ft
FLOW MODE ONLY
RELAYS:1 2 3 4 5 6
Tot:
EC 9