SENSORMOUNTING/LOCATION
-OpenChannelFlowApplications
Eachsensorisequippedwitha3/4inchisolationcouplingwhichMUSTbeusedinyourinstallation.A
threadednippleorlengthofconduitmaybeusedtopositionthesensoratthedesiredheight.The
sensorshouldbehand-tightenedbyturningthesensorstemonly.DONOTuseawrenchanddonot
overtighten.
IMPORTANT:Followtheflumemanufacturer’sdirectionsfortransducerlocation.Thetransducer
shouldbecenteredabovetheflumeapproachsectionandmounted12"/30.5cm(minimum)abovethe
maximumliquidlevel(dependingonsensormodel).
Thetransducershouldbeprotectedfromphysicaldamageandthetransducercableshouldberoutedin
aseparatemetalconduit.
Becausethesensorisequippedwithatemperaturesensor,itshouldbeshieldedfromdirectsunlight.
UsethePVC"isolationcoupling"suppliedwiththesensorand hand-tighten thesensor/coupling
assemblyontoyourmountingstand. Donot clampdirectlytothesensorortotheisolationcoupling.
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 21
12"
305
12"
305 mm