ENCLOSUREINSTALLATION
Locatetheenclosurewithin500ft(150m)ofthesensor.Itcanbewallmountedwithfourmounting
screws(supplied)orpanelmountedwithOptionPMPanelMountingKitfromGreylineInstruments.
Avoidmountingtheenclosureindirectsunlighttoprotecttheelectronicsfromdamagedueto
overheatingandcondensate.Inhighhumidityatmospheres,orwheretemperaturesfallbelowfreezing,
OptionTHEnclosureHeaterandThermostatisrecommended.Sealconduitentriestopreventmoisture
fromenteringenclosure.
NEMA4X(IP66)WITHCLEARCOVER
1.Openhingedenclosurecover.
2.Insert#8screwsandwashersthroughthefourenclosure
mountingholestosecureenclosuretowallormountingstand.
3.Closecover.
Additionalconduitholescanbecutintheendoftheenclosure
whenrequired.UseaholesaworGreenlee-typeholecuttertocut
therequiredholes.
Note:Thisnon-metallicenclosuredoesnotautomaticallyprovidegroundingbetweenconduit
connections.Groundingmustbeprovidedaspartoftheinstallation.Groundinaccordancewiththe
requirementsoftheNationalElectricalCode.Systemgroundingisprovidedbyconnectinggrounding
wiresfromallconduitentriestothesteelmountingplateoranotherpointwhichprovidescontinuity.
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 24
END VIEW
COVER
ENCLOSURE
ENCLOSURE
MOUNTING
HOLES