CALIBRATIONWORKSHEET-FlowMode
forOpenChannelFlowthroughFlumeandWeirs
MAXRANGE A=_______________
(SLT32READS)
MAXHEAD B=________________
(HeightatMAXFlow)
A - B = MINRANGE
______ - ______ = ____________
NOTE
:TheSLT32willnotdetecttargetsbeyondtheuserentered MaxRg.
Page 53
SLT32Level&FlowMonitor
ManualSeriesB.11
A
B
SENSOR
MINRANGE
MUST BE
12 in. (305 mm)
ORMORE
UNITS
SELECTION
ftinmcm%
RANGE
STORE?YES
Program SLT32toshow
RANGE via:
CALIBRATION
MODE
MINRANGE
MAXRANGE
STORE?YES
Program MINRANGE then
MAX RANGE via: