APPENDIXA-OPTIONS
EXTRASENSORCABLE
(OPTIONXC)
EachGreylineSLT32includes25ft.(7.6m)RG62AUcoaxialcable.AdditionalRG62AUcoaxialcable
andCableJunctionBox(OptionJB)maybeorderedwiththeFlowMonitor,orthecablemaybespliced
andextendedupto500ft(152m)asrequiredduringinstallation.Noadjustmentisrequiredwhenthe
sensorcableisextendedorshortened.UseonlyRG62AU(orRG62U)coaxialcablewhichisavailable
fromGreylineInstrumentsoryourlocaldistributor.NominalimpedanceofRG62AUcableis93ohms.
Extendedsensorcablemustbeinstalledinmetalconduittopreventinterference.DonotuseBNC
coaxialconnectors(TVcabletype).Recommendedinstallationwithametaljunctionboxisillustrated
below:
Note:OptionalWatertightsteelNEMA4JunctionBoxeswithterminalstrips(OptionJB)areavailable
fromGreylineInstruments.
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 34
SENSORCABLE
25 ft (7.6 m)
RG62AU COAXIAL