Greyline Instruments SLT32 - Wiring Problems and Non-Linearity

Greyline Instruments SLT32
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C -through4-20mAoutputcable -useshieldedtwistedpair(shieldedtoAC
ground)
-usegroundedmetalconduit
C -wiringorinstallationcloseto
variablespeeddriveorinverter
-followV.S.D.manufacturer’sinstructions
forDrivegrounding,wiringandshielding
UnitReceivesNoReturnEchoDueTo:
C,F,E -foamonliquidsurface -usestillingwell(openchannelflow)
B -targetbeyond MaxRg -recalibrate
E -calibrationerror - ISBNo
selectedin SPECIALFUNCTIONS
andinstrumenthasanoptionalISB
F -sensordamaged -removesensorfrommountingandaimata
flat,stabletargettotest
WiringProblemsDueToSensorCable:
A,C,F, -opencircuit -checkconnections/continuity(8850to
12700ohmsmax./-30°Cto+70°C)
B,F -shortcircuit -checkconnections/continuity(8850ohms
min.)
F -toolong(max500ft.,150m)
C -bundled/runinconduitwithpower
cable
C -sensorgroundshortedto
conduit/enclosure
-insulate
A -extendedwithwrongtypeofwire -useRG62A/Ucoaxialonly
C -closetohighvoltage/largemotors
C -ACchassis/groundmissingon
instrumentpowerconnections
Non-LinearityDueTo:
H -vapour -dissipatefumes,Calibrationin-situ
H -zeronotsetaccurately -see“ZeroPositioningofSensor”
H -wrongflume,orK&nselected
(FLOW mode)
-selectcorrectflume
D,E -calibrationerror - ISBYes
selectedin SPECIAL
FUNCTIONS menuandinstrumentdoesnot
haveanISB
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 27