POWERINPUTOPTION
12VDCOR24VDC
SLT32Level&FlowMonitorsmaybeorderedfactory-configuredfor12VDC,or24VDCpowerinput.
QUICKBENCHTEST:
ConnectSensorasshownbelow,thenPower.Whenproperlyconnectedasoftclickingcanbeheard
fromthesensorandfigureswillshowonthelargeLCDdisplay.TestoperationoftheSLT32byholding
thesensorsteadilyandaimingataflat,stabletarget12to28"(305to711mm)awayfromtheendofthe
sensor.AllowafewsecondsfortheSLT32tolockontothetargetbeforedisplayingitsdistance.The
SLT32willnowdisplayRangeinftorcm(factorycalibration).
CONNECTIONS:
POWERINPUT:Connectonly12VDC/0.5Ampstothe+and-terminalsforunitsmarked12V,or
24VDC/0.5Ampsforunitsmarked24VDC.ThePowerInputGNDmustbeconnectedtothenearest
Groundpole.A1ampfuseinlineisrecommended.PowerConsumptionis6.5Wcontinuous.
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 37
GND
12 AWG
MAX
OPTIONAL 24VDC
OPTIONAL 12VDC
POWERINPUT
IMPORTANT:
MUST CONNECT TO A
GOODGROUND (
<1 Ohm)
WITH12 AWG CONDUCTOR
24VDC
12V or
1 AMP FUSE