RS485SERIALOUTPUT
(ReplacesstandardRS232OutputwithDataLoggeroption.)
PermitstransferofoptionalDataLoggerreportsonwirelengthsupto4000ft.(1220m).
Format:8Bits,1StopBit,NoParity.(BaudratemaybechangedundertheSPECIALFUNCTIONS
menu.Factorydefaultis19200baud).
Useshielded,4-conductorcable(GreylinecableoptionSC-18AWGrecommended).Connectthecable
shieldtoGroundattheSLT32electronicsenclosureanddonotconnecttoGroundattheotherend.
ConfiguretheGreylinesoftwareprogramforRS485communications.UndertheCommunications/
ConnectionSetupmenu,select'ConnectedtoaGreyline485OI-12Vadaptor'.
485OI9TB-12V
RS232/RS485Converter
UsewithoptionalRS485outputforconnectiontoaPCcomputer
ormodem.
•
Opticallyisolatesandprotectsyourcomputer'sRS232port
•
TerminalblockforRS485connections
•
Operatesfrom2400upto19.2Kbaud
DB9M/DB9F-RS232Cable
ConnectsRS232/485ConvertertoaPC
Cablelength6ft.(2m)withDB9MandDB9Fconnectors.
SLT32Level&FlowMonitor
ManualSeriesB.11
Page 44
Greyline
Transmitter
with
RS485 Output
RS485 <4000 ft (1220 m)
Detail
RX
RS485
PU
GND
TX
(GREEN)
(WHT)
(BLK)
(RED)
18AWG, 4-conductor, shielded
Option: SC-18AWG
DB9F
DB9M
RS232 Cable
Option: DB9M/DB9F
6 ft (2 m)
RS232/485 Converter
Option: 485OI9TB-12V
RX
TX
GND
PU
RS485
RS232
DB9F