GDM-906X Series User Manual
346
Replace 3A Input Current Fuse
Preparation
To make sure if 3A input current needs to be replaced, press
the button to set GDM-9060/9061 in Continuity mode
and short circuit the HI input terminal with the 3A input
current terminal.
If the test result shows OPEN, either one of the fuses
requires replacement. The one is accessible from the rear
panel, and the other one is within the internal part.
If either one of the fuses of 3A input current is damaged,
please first check the one (3.15 A500 V) in the lower-left
corner of rear panel.
Step
1. Turn the GDM-9060/9061 off
2. Press and hold the fuse holder in the rear panel followed
by rotating it counterclockwise with a flat-blade
screwdriver.