GDM-906X Series User Manual
46
F3 (Auto Zero)
key to enable Auto
Zero
Background
Autozero provides the most
accurate measurements, but
requires additional time to
perform the zero measurement.
With autozero enabled (On), the
GDM-9060/9061 internally
measures the offset following each
measurement. It then subtracts
that measurement from the
preceding reading. This prevents
offset voltages present on the
GDM-9060/9061’s input circuitry
from affecting measurement
accuracy. With autozero disabled
(Off), the GDM-9060/9061
measures the offset once and
subtracts the offset from all
subsequent measurements.
Display
When turning On the Auto Zero, the display
shows an icon indicating the Auto
Zero mode is currently being activated.
F4 (BeepVol) key
to select the Vol
Press the F2 ~ F4 key to select the volume level
or press the F1 key to set Beep volume off
Set Continuity Threshold
Background
The continuity threshold defines the maximum resistance
allowed in the DUT when testing the continuity.
Threshold Range
Threshold
1 to 1000Ω (Default Threshold:10Ω)
Resolution
1Ω