REMOTE CONTROL
155
Example
:SEQUence2:ENDState LAST
Sets the terminate state of sequence as “the last
group”.
Command
:SEQUence[1|2]:ENDState?
Function
Queries the terminate state of the sequence.
Example
:SEQUence2:ENDState?
Queries the CH2terminate state of the sequence.
Returns to OFF or LAST.
Command
:SEQUence[1|2]:GROUPs <value>
Function
Sets the output group number of sequence.
Description
<value> the definition of output group number is
that the default voltage/current group number of
power output from each cycle ranges from 1 to
2048. All group number of sequence = output
group number*cycle number. Power will terminate
timer function after completing outputs of all
group numbers.
Example
:SEQUence2:GROUPs 25
Sets the CH2 output group number of sequence as
25.
Command
:SEQUence[1|2]:GROUPs?
Function
Queries the output group number of the sequence.
Example
:SEQUence2:GROUPs?
Queries the CH2 output group number of the
sequence. Returns to an integer between 1 to 2048.
Command
:SEQUence[1|2]:PARAmeter<No>,<volt>,<curr>,<
time>
Function
Sets the parameter of the specified group.