EasyManuals Logo

GW Instek GPP-2323 User Manual

GW Instek GPP-2323
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
SYSTEM SETTINGS
93
Firmware Upgrading
When to
Upgrade
Firmware
When system is failure, request by customer GW
Instek. When the system fails, firmware can
requested by GW Instek customers.
Upgrade
Requirement
Firmware file
Supplied by
GW Instek
Operation
After plugging in flash drive, the corresponding
symbol appears in the status bar.
Press System button;
Press F3 (Setting) button;
Press F4 (Upgrade) button;
Rotate the scroll wheel to select corresponding
file;
Press the F4 (Recall) button and start to upgrade;
The procedure is done when the message,
upgrading successful, shows up.
Note
Flash drive only support USB2.0/USB3.0,FAT32
file system,16G MAX.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GPP-2323 and is the answer not in the manual?

GW Instek GPP-2323 Specifications

General IconGeneral
BrandGW Instek
ModelGPP-2323
CategoryPower Supply
LanguageEnglish

Related product manuals