GPP Series User Manual
136
Description
[FAST]: In order to protect the function switching
device relay, it is necessary to wait until the voltage
and current drop to zero under normal conditions.
When FAST is enabled, this factor will not be
considered, but it is easy to damage the relay, so it is
recommended to use it with caution.
Example
:LOAD2:CC ON
Sets CH2 as Load CC mode.
Command
:LOAD[1|2]:CR {ON|OFF}
Function
Sets CH1/CH2 as Load CR mode.
Example
:LOAD2:CR ON
Sets CH2 as Load CR mode.
Command
:MODE[1|2]?
Function
Queries CH1 or CH2 work mode.
6 modes below:
SERiesPARallelINDE pendent
CV LoadCC LoadCR Load
Example
:MODE1?
Queries CH1 work mode.:
Command
:LOAD[1|2]: RESistor <NRf>
Function
Sets the Load CR level.
Description
<NRf>
1-1000
Example
:LOAD2: RESistor 100
Sets the Load CR level to 100Ωfor CH2.
Command
:LOAD[1|2]: RESistor?