PSP Series User Manual
50
SV
Command
The SV command sets the voltage output level.
SV <parameter> <CR>
ASCII
53 56 <parameter> 0D
HEX
<parameter>
Description
xx.xx
Voltage value( 5 characters, 2 decimal
places)
SV 10.50<CR>
Sets the voltage output to 10.5V.
SU
Command
The SU command sets the voltage output limit.
SU <parameter> <CR>
ASCII
53 55<parameter> 0D
HEX
<parameter>
Description
xx
Voltage value( 2 characters, no
decimal)
SU 11<CR>
Sets the voltage limit to 11V.
SI
Command
The SI command sets the current limit. The power
limit will be automatically altered to suit.
SI <parameter> <CR>
ASCII
53 49<parameter> 0D
HEX
<parameter>
Description
x.xx
Current limit value( 4 characters, two
decimal places)
SI 1.10<CR>
Sets the current limit to 1.10A.