PSU Series User Manual 
32 
PSU
Load
Ext-V
Ext-R
Analog 
connector
(      ) Insulation capacity ≥ isolation voltage 
of power supply
If the insulation capacity of the load and load cables 
are not greater than the isolation voltage of the power 
supply, electric shock may occur. 
If the positive or negative terminal is connected to 
the protective ground terminal, the insulation 
capacity needed for the load and load cables is 
greatly reduced. The insulation capacity only 
needs to be greater than the maximum output 
voltage of the power supply with respect to 
ground. 
PSU
Load
Ext-V
Ext-R
Analog 
connector
(      ) Insulation capacity ≥ voltage of power 
supply with respect to ground
If using external voltage control, do not ground the 
external voltage terminal as this will create a short 
circuit.