PSU Series User Manual 
94 
Configuration Overview 
Configuration of the PSU power supplies is divided into five 
different configuration settings: Normal Function, USB/GPIB, LAN, 
UART, System Configuration Settings, Power ON Configuration, 
Trigger Input/Output Configuration Settings and Special Function 
Settings. Power ON Configuration differs from the other settings in 
that the settings used with Power ON Configuration settings can 
only be set during power up. The other configuration settings can 
be changed when the unit is already on. This prevents some 
important configuration parameters from being changed 
inadvertently. Power On Configuration settings are numbered F-90 
to F-98 and the other configuration settings are numbered F-00 to F-
61, F-70 to F-78, F-88 to F-89 and F100 to F122. The Special Function 
Settings are used for calibration, firmware updated and other 
special functions; these functions are not supported for end-user 
use. 
Configuration Table 
Please use the configuration settings listed below when applying 
the configuration settings. 
 
V-I mode slew rate select  
0 = CV high speed priority (CVHS) 
1 = CC high speed priority (CCHS) 
2 = CV slew rate priority (CVLS) 
3 = CC slew rate priority (CVLS) 
0.001~0.060V/msec (PSU 6-200) 
0.001~0.080V/msec (PSU 8-180) 
0.001~0.125V/msec (PSU 12.5-120) 
0.001~0.150V/msec (PSU 15-100) 
0.001~0.200V/msec (PSU 20-76) 
0.001~0.300V/msec (PSU 30-50) 
0.001~0.400V/msec (PSU 40-38) 
0.001~0.500V/msec (PSU 50-30)