5.  Press the Voltage knob to save the 
configuration setting. ConF will be 
displayed when it is configuring. 
6.  If CC Slew Rate Priority was chosen as the 
operating mode, set F-06 (Current Slew Rate 
Up) and F-07 (Current Slew Rate Down) and 
save. 
0.001A~2.000A / msec (PSU 6-200) 
0.001A~1.800A / msec (PSU 8-180) 
0.001A~1.200A / msec (PSU 12.5-120) 
0.001A~1.000A / msec (PSU 15-100) 
0.001A~0.760A / msec (PSU 20-76) 
0.001A~0.500A / msec (PSU 30-50)   
0.001A~0.380A / msec (PSU 40-38) 
0.001A~0.300A / msec (PSU 50-30) 
0.001A~0.250A / msec (PSU 60-25) 
0.001A~0.190A / msec (PSU 80-19) 
0.001A~0.150A / msec (PSU 100-15) 
0.001A~0.100A / msec (PSU 150-10) 
0.001A~0.025A / msec (PSU 300-5) 
0.001A~0.008A / msec (PSU 400-3.8) 
0.001A~0.006A / msec (PSU 600-2.6) 
7.  Press the Function key again to exit 
the configuration settings. The 
Function key light will turn off.