11 Technical data
11.1 Running machines, 150/50 LC (pluto
®)
pluto
®
pluto
®
pluto
®
lt
pluto
®
med
pluto
®
lt med
UserTerminal
(display & keyboard)
6 LCD displays / 6 keys / display of mode and unit with LED.
“lt-models“ have no UserTerminal (no display, no keyboard). Control via interface.
Power supply AC
(standard, read nameplate),
dedicated line required.
U: 200 V ~ to 240 V ~
including N and PE / f: 50/60 Hz
Fuse (standard, read nameplate)
I
N
16 A at 230 V ~ (thermal fuse protected)
Power input (long time)
1150 VA
Power input (momentary)
3640 VA
(peak performances for milliseconds; max. 1s/min during heel strike)
Drive motor capacity
2200 VA
Elevation motor capacity
480 VA
Safety standards
c
IEC 60335-1 (VDE 0700)
IEC 60601-1-2,
VDE 0701,
ISO 20957-1,
EN 957-6,
n
MDD, directive 93/42/EEC + 2007/47/EC,
IEC 60601-1 (VDE 0750),
VDE 0751,
IEC 60601-1-2,
ISO 20957-1, EN 957-6,
Earth leakage current
1.0 mA
0.12 mA
Touch leakage current
n.a.
Isolation transformer
---
1840 VA
Degree of protection:
class I M / --- / IP 20
class I M / type B R / IP 20
Overvoltage category
Life-Earth < 300 V
mains transient voltage peak 1500V
Category II: Life-Earth < 300 V
transient over-voltage limit 2500 V
Mode of operation
continuous operation with intermittent loading (according to IEC 60601-1)
Field of application / accuracy
sports and fitness: S,I / A (ISO 20957-1; EN 957-6)
sports and medical: S,I / A (ISO 20957-1; EN 957-6)
Classification according to MDD
---
active therapeutic device and active diagnostic device,
risk class II b
Speed
Definition
0.5 ... 18.0 km/h
0.1 km/h
Speed accuracy*
+/- 5 %
Angle of inclination**
Definition
0.0 ... 20.0 %
0.1 %