11.3 Running machines, 170-190/65 (quasar / pulsar)
stellar
®
, quasar
®
, pulsar
®
locomotion
®
190/65 med
6 LCD displays / 6 keys / display of mode and unit with LED.
“lt-models“ have no UserTerminal (no display, no keyboard). Control via interface.
E: no Terminal
DE: TouchScreen
(standard, read nameplate),
dedicated line required.
U: 100 V ~ / 110 V ~ / 200 V ~ / 208 V ~ / 230 V ~
including N and PE
f: 50/60 Hz
Fuse (standard, read nameplate)
(peak performances for milliseconds; max. 1s/min during heel strike)
IEC EN 60335-1(VDE 0700)
IEC EN 60601-1-2,
VDE 0701, ISO 20957-1, EN 957-6,
Machinery Directive 2006/42/EC
MDD, directive 93/42/EEC + 2007/47/EC,
IEC EN 60601-1 (VDE 0750), VDE 0751,
IEC EN 60601-1-2, ISO 20957-1, EN 957-6,
Machinery Directive 2006/42/EC
class I M / type B R / IP 20
transient over-voltage limit 1500 V
Category II: Life-Earth < 300 V
transient over-voltage limit 2500 V
continuous operation with intermittent loading (according to IEC 60601-1)
Field of application / accuracy
sports and fitness: S,I,A (EN957)
sports and medical: S,I,A (EN957)
Classification according to MDD
active therapeutic device and active diagnostic device, risk class
II b
0.0 ... 10.0 km/h, 0.0 ... 40.0 km/h, 0.0... 45.0 km/h