technical data
file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 30.01.2015 printed 30.01.2015 page: 168 of 216
11.5 Running machines, oversize 200-450/75-300 (venus / saturn)
venus 200/75
venus 200/75 r
venus 200/100
venus 200/100 r
saturn 250/75
saturn 250/75 r
saturn 250/100
saturn 250/100 r
saturn 250/125 r
saturn 300/75
saturn 300/75 r
saturn 300/100
saturn 300/100
r
saturn 300/125 r
saturn 450/300 rs
External control unit /
UserTerminal (display &
keyboard)
External control unit with TouchScreen
Power supply AC
(standard, read nameplate),
dedicated line required.
200 V 3N~ / 208 V 3N~ / 400 V 3N~
with N and PE
(peak performances for milliseconds; max. 1s/min during heel strike)
Fuse (standard, read nameplate)
3 x C 32 A at 400 V 3N~
3 x 63 A
3 x 50 A
1 x 16 A
Power input (long time)
7500 VA
41500 VA
Power input (momentary)
15200 VA
66500 VA
Drive motor capacity
11000 VA
30000 VA
Elevation motor capacity
370 VA
18000 VA
Safety standards
n
MDD, directive 93/42/EEC + 2007/47/EC,
IEC EN 60601-1 (VDE 0750), IEC EN 60601-1-2,
VDE 0751, ISO 20957-1, EN 957-6, Machinery Directive 2006/42/EC
Earth leakage current
0.25 mA
on request
Isolation transformer
3 x 3000 VA
3 x 21000
VA
Degree of protection
class I M / type B R / IP 20
Overvoltage category
Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt
Mode of operation
continuous operation with intermittent loading (according to IEC 60601-1)
Field of application / accuracy
according to EN 957
sports and medical / S,I,A (EN957)
[r] = suitable for bicycle and wheelchair application (from running surface 3000 x 1250 mm standard)
Classification according to MDD
active therapeutic device and active diagnostic device, risk class II b
Speed / definition
0.0 ... 40.0 km/h / 0.1 km/h
0.0 ... 11.11 m/s / 0.1 m/s
Speed accuracy*
+/- 5 % (class A based on EN 957)
Option special speed
0.0 ... 30.0 or ... 50.0 or ... 60.0 or ... 80.0 km/h
on request higher speed versions than 40 km/h only for sports models
Angle of inclination**
Definition
-35 % ... +35 %
0.1 %
-27 % ... +27%**
0.1 %
-5 %
(-25 %) …