5.4.4 Switch off procedure model saturn 450/300
Press the “power off” push button.
Switch off the running machine by using the main
switch at the right side of the external UserTerminal
(control unit). The main switch has to be turned 90°
counterclockwise as shown in the picture.
The main switch can be secured against
unauthorized access with a padlock.
N The ON-/OFF intervals must not be shorter than 1-2 minutes (for models with 3-phase
connection: 2-3 minutes). This may otherwise lead to interference and/or failure of the backup
and/or data loss and/or data corruption.
N Models for medical application with the isolation transformer have an inrush current limiter
(surge guard). Too short ON/OFF intervals will lead to a deactivation of the inrush current limiter
and result in an overload of the circuit fuse.
N If the device is used daily, we recommend switching it on in the morning and leave it in stand-by
mode during the day.
N Press the emergency stop only in case of danger.
N The emergency stop is not to be used as a normal stop key.
5.5 Emergency Stop, stop buttons, stop functions and safety features
In case of emergency, e.g. danger of stumbling and/or falling, etc., grab the front crossbar and/or both side handrails
and jump onto the foot rails (step platforms) with both feet.
The emergency stop button is to be pressed immediately!
h/p/cosmos is in compliance with the norm EN 957-6 and offers more safety features than required, as shown in the
following tables. In all below mentioned cases, all displayed values will be reset to “0”.
N Use the emergency stop only if in danger.
N The emergency stop is not to be used as a normal stop key.