file: n:\article\cos14310m5-v1_08hpc-en\20150130_cos14310m5-v1.08hpc-en_instructions_for_use_h-p-cosmos_treadmill.doc
© 2015 h/p/cosmos sports & medical gmbh author: fh email@h-p-cosmos.com created 30.01.2015 printed 30.01.2015 page: 77 of 216
List of user options/optional functions
Reset (deleting) of error messages
Note: This option resets exclusively
the causing variable. If for example
the E.02 service-time-interval is
being set back with this operation,
the distance-interval is not set back
and vice versa! Only administrator
option OP 47 resets all three values
simultaneously.
The required maintenance work has to be performed before deleting
error messages. The h/p/cosmos service department has to be
contacted before one of the service displays or error codes will be
deleted!
Confirmation on display § indicating "donE"
Info: This option only sets back the error message! If the error still
exists, you are not able to set back the error message. In this case
consult an authorised service engineer
The following intervals are set back by this operation:
E.01: Oil-interval A-OP 35
E.02: Service-time-interval A-OP 37
E.02: Service-distance-interval A-OP 38
Total distance covered (km)
§ and $ indicates: total distance covered in km
Indication of total hours of operation
= stand-by-time including runtime of
motor/ running belt (h)
§ and $ show: operation hours
Indication of total hours of
runtime of motor/running belt (h)
§ and $ report: operation hours
Indication of firmware version and
date
= reports “OP05” ! reports “typE”
" indicates device type, e.g. “1.4”
§ indicates: „MCU 5“ $ reports version, e.g. “1.01.1”
% reports default type, e.g. 1.3
Adjustment of actual date and the
real time clock
! shows: rtc for Real Time Clock
$ reports flashing: date / time, year, month, date, hours, minutes,
seconds
Acoustic heart rate signal
This function is normally used to control the regularity of the heart rate
or to find reasons for transmitting problems
§ indicates: OFF or ON
OFF: no acoustic heart-rate-signal
ON: acoustic heart rate signal for every beat
Stop time / deceleration time after
STOP key in relation to the max.
speed
§ indicates: stop time in sec.
$ indicates: sec for seconds
Adjustable from 2.... 30 seconds