25
SECTION 2, continued
1. Fill the clean sample
cell to the line with
high quality water
(< 0.5 NTU). Cap and
wipe with lint-free cloth.
Apply silicone oil. See
Section 3.6.2.2 on
page 40 for more
information about
high quality water.
2. Press: I/O to turn the
instrument on.
3. Insert the sample
cell into the sample
compartment. Close
the cover.
4. Press: READ
Record the cell's position
in the cell compartment
and the displayed
reading.
Note: This procedure may
be easier if the user holds
the
READ key through the
whole process. This allows
the lamp to remain on and
make continuous readings.
5. Remove the cell,
rotate it slightly and
reinsert it into the cell
compartment. Close the
cover, then press
READ.
Record the cell's position
and the displayed
reading.
6. Repeat step 5 until
the lowest reading is
displayed. Place an
orientation mark on the
cell's marking band near
the top of the cell so the
cell can be consistently
inserted in the position
that yields the lowest
reading. When using the
cell, always place it in
the instrument so the
orientation mark aligns
with the raised mark on
the instrument.
I
O
READ

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P and is the answer not in the manual?

Hach 2100P Specifications

General IconGeneral
Range0 to 1000 NTU
Light SourceTungsten filament lamp
DetectorSilicon photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Power4 AA alkaline batteries
Resolution0.01 NTU
Measurement PrincipleNephelometric
DisplayLCD

Related product manuals