SECTION 2, continued
7. Press: READ
Record the cell’s position
in the cell compartment
and the displayed
reading.
8. Remove the cell,
rotate it slightly and
reinsert into the cell
compartment. Close the
cover, then press
READ
again. Record the cell’s
position and the
displayed reading.
9. Repeat step 8 until
the value displayed for
the second cell is within
0.01 NTU (or 1%) of the
value obtained for the
first cell. Place an
orientation mark on the
second cell's marking
band so it is consistently
inserted in this position.
Note: Due to variability
in glass, it may not be
possible to match all cells.
10. Repeat step 6
through step 9 if
matching other
sample cells.
READ

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100P and is the answer not in the manual?

Hach 2100P Specifications

General IconGeneral
Range0 to 1000 NTU
Light SourceTungsten filament lamp
DetectorSilicon photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Power4 AA alkaline batteries
Resolution0.01 NTU
Measurement PrincipleNephelometric
DisplayLCD