EasyManua.ls Logo

Hach DR2400 - Page 201

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DR/2400
Chlorine, Total
ChlorineTot_PT_ULow_DPD_Eng_Ody.fm Page 1 of 8
• Analyze samples immediately. Chlorine is not stable in an aqueous solution.
• A reagent blank value for a combined lot of indicator/buffer reagent solutions should be determined at least once a day. If sample
color or turbidity fluctuates frequently during the day, determine a reagent blank for each sample. See Determining the Reagent
Blank Value on page 3. To store the reagent blank value, see Creating a User Program Based on a Hach Program in the
instrument manual.
• The ampules contain more than 1.0 mL of solution for ease of transfer. Discard excess reagent in the ampule.
• Install the Pour-Thru Cell into the adapter so that the long pathlength (1-in./25-mm) is parallel to the front of the instrument.
Chlorine, Total
(Method 8370
DPD Method*
* U.S. Patent 5,362,650
Pour-Thru Cell
ULR (2 to 500 µg/L as Cl
2
)
Scope and Application: For detecting trace levels of chlorine and chloramines in clean waters relatively
free of color and turbidity; USEPA accepted for reporting for drinking water analysis
1. Touch
Hach Programs.
Select program
86 Chlor. Total ULR.
Touch
Start.
2. Install the Pour-Thru
Cell and the multi-
pathlength cell adapter in
the sample cell
compartment. Use the 25-
mm cell pathlength.
Note: See Section 2.6.1 on
page 26 in the Instrument
manual for installation details.
Flush with 50-mL of
deionized water.
3. Pour at least 50 mL of
sample into the
Pour-Thru Cell.
4. After the flow stops,
touch the timer icon.
Touch
OK.
A three-minute reaction
period will begin. This
time allows turbidity or
solids to settle and ensures
a stable reading.
Tips and Techniques
DR/2400 Pour-Thru Cell Method 8370
Hach Programs

Table of Contents

Related product manuals