DR/2400
Oxygen Demand, Chemical
OxygenCOD_Other_Other_MD3_Eng_Ody.fm Page 1 of 8
• Use this method if the sample contains chloride. To determine if the sample contains chloride, use Quantab
®
Titrator Strips for low
range chloride (Cat. No. 27449-40).
• If the sample COD is expected to exceed 1000 mg/L, dilute the sample as described inTa bl e 1 on page 2.
• Blending the sample promotes even distribution of solids and improves accuracy and reliability.
• Run one blank with each lot of reagent. Run all samples and blanks with the same lot of vials. The lot number appears on the
container label.
• The stability of the reagent blank allows for reuse. Verify the reagent blank quality by measuring the absorbance of the blank vs. a
clean COD vial filled with deionized water. The absorbance range should be about 1.41–1.47.
• If the sample boils during the digestion, the vial is not properly sealed. Test results will be invalid.
• Place a safety shield in front of the COD reactor to prevent injury if splattering occurs.
• Spilled reagent will affect test accuracy and is hazardous. Do not run tests with spilled vials.
Oxygen Demand, Chemical
Method 10067
Manganese III Reactor Digestion Method
(with optional chloride removal)*
* U.S. Patent 5,556,787
(30 to 1000 mg/L COD Mn)
Scope and Application: For water and wastewater
Note: Mixing concentrated
sulfuric acid and water is not
additive. Adding 1.0 mL of
concentrated sulfuric acid to
9.0 mL of sample does not
result in a final volume of
10.0 mL. This factor is built into
the calibration curve.
1. Turn on the COD
Reactor and heat to
150 °C. Place the safety
shield in front of the
reactor.
2. Homogenize 100 mL
of sample for 30 seconds
in a blender.
If suspended solids are
present, continue to mix
the sample while
pipetting.
3. Pipet 9.0 mL of
homogenized sample
into an empty glass
mixing cell (the prepared
sample) and 9.0 mL of
deionized water into
another empty glass
mixing cell (the blank).
Tips and Techniques for Preparing the Acidified Sample
Preparing the Acidified Sample

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach DR2400 and is the answer not in the manual?

Hach DR2400 Specifications

General IconGeneral
BrandHach
ModelDR2400
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals