EasyManua.ls Logo

Hach DR2400 - Page 620

Hach DR2400
798 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Phosphorus, Reactive (Orthophosphate)
Phosphorus, Reactive (Orthophosphate)
Page 2 of 6
PhosphorusReac_None_Other_AMI_Eng_Ody.fm
5. Touch the timer icon.
Touch
OK.
A 10-minute reaction
period will begin.
Continue with step 6 while
the timer is running.
6. Pour 25 mL of sample
into a round sample cell
(this is the blank).
7. When the timer
beeps, place the blank
into the cell holder.
8. Touch Zero.
The display will show:
0.00 mg/L PO
4
3–
9. Pour the prepared
sample into a round
sample cell.
10. Place the prepared
sample into the cell
holder.
11. Touch Read.
Results will appear in
mg/L PO
4
3–
.
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
Zero
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
Read

Table of Contents