DR/2400
Volatile Acids
VolatileAci_None_Other_ESR_Eng_Ody.fm Page 1 of 6
Volatile Acids
Method 8196
Esterification Method*
*Adapted from The Analyst, 87, 949 (1962)
(27 to 2800 mg/L)
Scope and Application: For digestor sludges
1. Touch
Hach Programs.
Select program
770 Volatile Acids.
Touch
Start.
2. Pipet 0.5 mL of
deionized water into a
dry 25-mL sample cell
(the blank).
3. Filter or centrifuge
25 mL of sample.
Centrifuging is faster
than filtration.
4. Pipet 0.5 mL of the
filtrate or supernatant
into a second dry 25-mL
sample cell (the prepared
sample).
5. Pipet 1.5 mL of
ethylene glycol into each
sample cell. Swirl to mix.
6. Pipet 0.2 mL of 19.2 N
Sulfuric Acid Standard
Solution into each cell.
Swirl to mix.
7. Place both cells into
a boiling water bath.
Alternatively, the cells
may be boiled in a 500-mL
beaker.
8. Touch the timer icon.
Touch
OK.
A three-minute reaction
period will begin.
10 mL
20 mL
25 mL
Method 8196
Hach Programs
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL
10 mL
20 mL
25 mL