Manual operation
Use the setup utility software with a direct PC connection or over a LAN
connection to manually operate the
instrument.
1. Open the SetupUtility.exe file to start the setup utility software.
2. Select the Data Display tab. Refer to Figure 15 for the data shown.
3. Use the buttons to operate the instrument.
Note: The buttons change depending on the system status.
Option Description
Monitor Shows the updated data continuously in real time.
Stop Monitor Stops changes to the data shown.
Sample Starts sample collection. Samples are taken according
to the settings on the Basic Setup tab. Changes the
instrument to active mode if in inactive mode.
Stop Count Stops sample collection.
Active Mode Sets the internal laser to on. Enables alarms.
Inactive Mode Sets the internal laser to off. Disables alarms.
Display Buffered
Data
Shows the data from the last sample completed.
Updates as each sample is completed.
Download Buffer Saves a copy of the data records in the buffer to the
PC as a text (CSV) file.
Erase Buffer Erases all data records from the buffer.
Figure 15 Real-Time data display
1 Active alarms (Sensor, Flow,
Comm, Count
1
)
7 Air flow (cfm or L/min)
2 Channel sizes and particle counts 8 Air volume collected for the sample
(cfm or L/min)
3 Sample time 9 Values from the optional relative
humidity (RH) and temperature
sensor
4 Last sample start time and date 10 Not available
5 System status 11 Cal voltage—identifies the
cleanliness of the sensor optics
6 Service use only
1
When a count alarm occurs, the high particle count is shown in red.
22 English

Table of Contents