Table of Contents
1 Specifications on page 3
2 Online user manual on page 4
3 General information on page 4
4 Installation on page 5
5 User interface on page 16
6 Start up on page 17
7 Operation on page 17
8 Maintenance on page 26
Section 1 Specifications
Specifications are subject to change without notice.
Specification Details
Ambient temperature –5 to 50°C (23 to 122°F)
Storage temperature –20 to 70°C (–4 to 158°F)
Operating humidity 0 to 95% non-condensing relative humidity
Operating altitude 2000 m (6561 ft) maximum
Environmental
conditions
Indoor use
EMC requirements
EN61326-1: EMC Directive
Note: The wall mount instrument is a Class A product. In a domestic environment this product may
cause radio interference in which case the user may be required to take adequate measures.
Korean registration
User Guidance for EMC Class A Equipment
업무용을 위한 EMC 등급 A 장치에 대한
사용자 지침
사용자안내문
A 급 기기 ( 업무용 방송통신기자재 )
이 기기는 업무용 (A 급 ) 전자파적합기기로서 판매자 또는 사용자는 이 점을 주의하시기
바라며 , 가정외의 지역에서 사용하는 것을 목적으로 합니다.
Compliance
CE, ETL certified to UL and CSA safety standards (with all sensor types), FCC, KC,
RCM, EAC, UKCA, SABS
Pollution degree 2
Enclosure ratings
IP65; Totally protected against dust; Protected against low pressure jets of water from all
directions.
NEMA 4X (wall mount only); Totally protected against dust; Protected against pressure
jets of water from all directions.
W A R N I N G
Enclosure rating does not apply to external power supply for benchtop instruments.
Power supply Universal 100 VAC to 240 VAC ±10% @ 50/60Hz - 40VA; 10 to 30 VDC - 30W
Analog current output
version on the
measurement board
4-20 mA (default) or 0-20 mA (configuration with software);
3 configurable outputs; Maximum load: 500 ohm; Sensitivity: 20μA;
Accuracy: ± 0.5% (between operating temperature limits)
Analog voltage output
version on the
measurement board
0- 5 V output (hardware option);
3 configurable outputs; Minimum load: 10 KOhm; Sensitivity: 5 mV;
Accuracy: ± 0.5% (between operating temperature limits)
English 3