Transmitter 9125 – Conductivity / resistivity measurement
Page 31
Type of calibration
x With a 2 electrode probe
For the first point, remove the probe from liquid or
unscrew the connector from the probe.
For the second point connect a resistance to the
IN/OUT terminal of the conductivity module.
x With an inductive probe
For the first point, remove the probe from liquid.
For the second point connect a resistance with a
looped wire through the probe.