Transmitter 9125 – Conductivity / resistivity measurement
Page 36
PARAMETERS Menu
This menu displays the calibration parameters of the
conductivity measurement.
Date of last calibration. This date is the date the user
has entered after a conductivity calibration ( electric, 2
points, 1 point).
The other parameters are those which are displayed
when the date is registered :
x Slope of the last calibration which is a correction
factor of the conductivity probe slope.
x Si la mesure est effectuée en TDS, le coefficient
de proportionnalité entre la conductivité et la
concentration.
x Drift of the temperature measurement.
HISTORIC Menu
This menu displays the conductivity slopes
corresponding to the last two conductivity calibrations
and allows to follow the probe clogging.