Transmitter 9125 – Conductivity / resistivity measurement
Page 43
S/ALARMS menu
This sub-menu allows to reach the configuration of
alarms 1 to 4.
The MODE parameter allows to choose the operating
mode of the 4 alarms :
- Limit : alarms 1...4
- USP : only alarms 1 and 2
- System : only alarm 3
- Timer : only alarm 4
MODE : x No
x Limit
x USP
x System
x Timer
ALARM 1
ALARM 2
ALARM 3
ALARM 4
ALARMS
Select
MODE : Syst
ACCEPT : Manual
RELAY : NO
ALARM 3
Select
System
MODE : Timer
INTERV : 2440 mn
Impul. Nb : 1
Ton : 005 s
Toff : 005 s
TmA : 20 mn
ALARM 4
Select
Timer