Transmitter 9125 – Conductivity / resistivity measurement
Page 61
7. Error messages
Note :
In manual acquittal, in order to suppress an error
message press ENTER after correcting the default.
ERROR MESSAGE
DESCRIPTION/POSSIBLE CAUSE
¾ Sensor not correctly connected
Temperature sensor damaged
Replace it if necessary.
¾ Sensor not correctly connected
Temperature sensor damaged
Replace it if necessary.
¾ The resistivity value is inferior to the lower limit of the
measuring range
.
¾ The conductivity value is superior to the upper limit of
the measuring range
.
¾ The electrode polarization is too high.
10.3
mS/cm
Pt100/Pt1000
SHORT CIRCUIT
12.7
mS/cm
Pt100/Pt1000
OPEN CIRCUIT
60
:
.cm
MEASURE
TOO LOW
15.2
mS/cm
MEASURE
TOO HIGH
15.2
mS/cm
POLARIZATION
TOO HIGH