Transmitter 9125 – Conductivity / resistivity measurement
Page 35
Temperature electric calibration / Resistor adjust
This calibration is factory realised.
Used to realise an electric calibration for the Pt100
measurement. 2 resistances of known values should
be connected to the temp + and temp - of the module
measure. This resistors should have a precision about
0,1%.
Proceed as follows :
x Launch the calibration when you have configured
an electric calibration. Lancer l'exécution après
avoir configuré l'étalonnage en étalonnage
électrique.
x The transmitter requires the connecting of the first
resistance.
x Press Enter.
x Repeat the same procedure for the second
resistance.
Process calibration
Proceed as follows after you have configured a
process calibration :
x Wait till the measurement is stable and press the
right function key OK.
x You have the possibility to change the value
x Press Enter.
x The instrument executes a zero adjustment
required to display the value configured.