Transmitter 9125 – Conductivity / resistivity measurement
Page 44
ALARM 1/2 (limit)
AFFECT. : choice of a limit on the measurement
or on the temperature.
x No
x S/:
x °C/°F
LIMIT : value of the limit.
DIR. : choice of the direction :
x Up
x Down
DELAY : definition of the temporisation when
the relay is interlocking (in seconds).
HYST : definition of the hysteresis in % (10%
max).
RELAY : choice of the relays normally opened
or closed.
x N
O
x NF
ALARM 1/2 (USP)
The TEMP menu allows the temperature to be
entered in cases where no measurement is
provided by Pt100/Pt1000. If temperature is
measured automatically, this menu is not
displayed.
The LEVEL menu is used to set the safety margin
with respect to the USP standard: with a level of
50%, the 1.9 μS/cm threshold (50°C) drops to 0.95
μS/cm.