Transmitter 9125 – Conductivity / resistivity measurement
Page 62
ERROR MESSAGES DURING A CALIBRATION
Note :
Press ESC to leave the menu and calibrate again.
¾ The electronic zero shift is superior to the limit
programmed.
Calibrate again.
¾ The slope shift is superior to the limit
programmed.
Limits : 50...150 %
¾ The temperature drift is superior to the limit
programmed
.
Limits : r 20 °C
OFFSET
OUT OF LIMITS
SLOPE : 162 %
SLOPE
OUT OF LIMITS
'
T : -30.0°C
'
T
OUT OF LIMITS
COND. CALIB.
COND. CALIB.
TEMP. CALIB.