Loading...
Transmitter 9125 – Conductivity / resistivity measurement
Page 7
Connections
2,5 mm
2
screw terminals
Fuse
5 x 20 mm Cartridge
T2AL - 250 V
Consumption
25 VA
European standards
EN 61326-1997 and EN61326 A1-1998 (Industrial level
for immunity)
EN- 61010-1
UL and CSA agreement
File E226594
MECHANICAL CHARACTERISTICS
Dimensions
144 x 144 x 150 mm
Weight
2 Kg
Material
Housing : Polyester coated aluminium
Screws : stainless steel
Tightness
IP65
Mounting types
Wall
Pipe
Panel
Cable glands
2 x PG13.5
2 x PG11
PERFORMANCE
2 electrode probe
Cell constant
0,01 cm
-1
0,1 cm
-1
1 cm
-1
Conductivity range
0,01 PS/cm...200 PS/cm
0,10 PS/cm...2 mS/cm
1 PS/cm....20 mS/cm
2 electrode probe
Cell constant
0,01 cm
-1
0,1 cm
-1
1 cm
-1
Resistivity range
5 k:.cm...100 M:.cm
0,5 k:.cm...10 M:.cm
50 :.cm...1 M:.cm
Inductive probe
Cell constant
1 cm
-1
2,35 cm
-1
10 cm
-1
Conductivity range
100 PS/cm...1 S/cm
200 PS/cm...2 S/cm
1 mS/cm...10 S/cm

Table of Contents